GE Innova IGS 6 Preinstallation Manual page 112

Table of Contents

Advertisement

Figure 2-56 Fixing Bolt Overview
NOTE
For more details on Table Baseplate, refer to
on page
103.
NOTE
With any kind of fixation methods (Bolts M20, Mechanical anchors or Chemical anchors),
the number of holes used mandatory is:
• Gantry baseplate : 12 max and 8 min holes used are acceptable
• Table baseplate : 10 max and 8 min holes used are acceptable
• Floor baseplate : 24 max and 12 min holes used are acceptable
We can have only 2 consecutive holes omitted.
5750182-1EN 3
Figure 2-57 Gantry and table mounting holes
TM
Innova
IGS 6
Equipment Requirements
101