WARNING
PORTABLE RF COMMUNICATIONS EQUIPMENT INCLUDING PERIPHERALS (SUCH AS
ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER
THAN 30CM (12 INCHES) TO ANY PART OF THE SYSTEM INCLUDING CABLES
SPECIFIED BY THE MANUFACTURER.
OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS EQUIPMENT COULD
RESULT.
Recommended Separation Distances for Portable and Mobile RF
Communications Equipment IEC 60601-1-2 (Ed2.1&3)
Table 3-6
Frequency of Transmitter
Equation
Rated Power of Trans-
mitter (watts)
0.01
0.1
1
10
100
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the equation in
the corresponding column, where P is the power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE
These are guidelines. Actual conditions may vary.
Limitations Management
Adhering to the distance separation recommended in
2.5 GHz, will reduce disturbances recorded at the image level, but may not eliminate all disturbances.
However, when installed and operated as specified herein, the Innova system will maintain its essential
performance by continuing to acquire, display, and store diagnostic quality images safely.
For example, a 1 W mobile phone (800 MHz to 2.5 GHz carrier frequency) shall be put 2.3 meters (see (*)
Table 3-6 on page
125) apart from the Innova system (in order to avoid images interferences risks).
5750182-1EN 3
150 KHz to 80 MHz
d = [3.5 / V
]√P
1
Distance (meters)
0.11
0.37
1.1
3.7
12
TM
Innova
IGS 6
80 MHz to 800 MHz
d = [3.5 / E
]√P
1
Distance (meters)
0.11
0.37
1.1
3.7
12
Table 3-6 on page
125, between 150 kHz and
Special Construction Requirements
800 MHz to 2.5 GHz
d = [7 / E
]√P
1
Distance (meters)
0.22
0.74
2.3 (*)
7.4
23
125