GE Innova IGS 6 Preinstallation Manual page 99

Table of Contents

Advertisement

Equipment Requirements
Figure 2-46 Table baseplate
NOTE
Prepare the floor such that the Table baseplate will be flush with the floor surface.
For alternative table bolts or seismic area, refer to template drawing shown in Illustration
Gantry and table mounting holes contained in
Hole dimension and preferred location in concrete floor
In the examination room, the Innova Frontal Positioner is not placed on a computer floor but directly put
on concrete floor, the location of the cable access needs to be carefully planned.
Otherwise, if the cable run is located under the concrete floor, the cables will have to come through the
floor and in this case you will need two holes, one for the Frontal Positioner and the other for the patient
table.
The diameter of both holes must be the same 225 mm or 9 in.
88
2.3.2 Mounting Requirements on page
TM
Innova
IGS 6
92.