Cleaning And Disinfection Instructions - Hettich MIKRO 220 Operating Instructions Manual

Rotors and accessories
Hide thumbs Also See for MIKRO 220:
Table of Contents

Advertisement

Cleaning and care
Chap.
Task to execute
8.3
Cleaning the accessories
8.4
Disinfection
8.4
Disinfecting the device
8.4
Disinfecting the accessories
8.5
Maintenance
8.5
Greasing the rubber seal of the centrifuging
chamber
8.5
Greasing the rubber seal of the biosafety system
8.5
Checking the accessories
8.5
Checking the biosafety system
8.5
Inspecting the centrifuging chamber for damage
8.5
Greasing the motor shaft
8.5
Accessories with a limited service life
8.5
Replacing centrifuge tubes
8.2

Cleaning and disinfection instructions

32 / 40
DANGER
Risk of contamination for the user due to inadequate
cleaning or failure to observe the cleaning instructions.
Observe cleaning instructions.
Wear personal protective equipment when cleaning the
device.
Observe laboratory regulations (e.g. TRBAs, the German
Protection against Infection Act, hygiene plan) for han-
dling biological agents.
The device and its accessories must not be cleaned in dishwashers.
Only perform hand cleaning and liquid disinfection.
The water temperature must not exceed 25 °C.
To prevent any corrosion due to use of detergents or disinfectants, it is
essential to follow the special application instructions provided by the
manufacturers of the detergent or disinfectant.
Disinfectant:
Surface disinfectant (not disinfectant for hands or instruments)
Ethanol as the sole active substance.
Do not use an ethanol-propanol mixture to disinfect the viewing window
in the lid of the device.
Rev.: 04 / 11.2023
 
 