Malfunctions

10 Malfunctions
10.1

Malfunctions on the Operating Display or During Weighing Processes

Fault
Cause
The operating
The device is disconnected.
display is blank.
The displayed
The device installation site is unstable.
weight value
changes constantly.
A foreign object is positioned between
the weighing pan and the housing.
The weight readout
The device has not been adjusted.
displayed by the
The device was not tared before
device is obviously
weighing.
wrong.
The side or upper
The operating range of the side or upper
panel of the
panel of the motorized draft shield is
motorized draft
not functional or is blocked.
shield does not
open or close.
74
Operating Instructions Cubis
| MCA Models
®
Solution
Check whether the power
supply unit is connected to the
device and the power supply at
the installation site.
If required: Connect the
AC adapter.
Adjust the parameters in the
"Ambient conditions"
submenu.
Change the installation site.
Remove the foreign object.
Calibrate the device.
Tare the device.
Hold down the palm-operated
key. If the problem continues
to occur: Contact Sartorius
Service.
Chapter, Page
6.3, 48;
6.4, 49