Section 7 Troubleshooting
7-2
Warning Displays
NOTE The heating element and pump continue to run with a warning.
A warning also sounds the alarm, if enabled.
Low Level Warn.
please remove reason
and press ENTER
to clear message
High Level Warn.
please remove reason
and press ENTER
to clear message
Other Errors
A/D Converter FAULT-
please remove reason
and press ENTER
to clear message
Bath Communication Error
please remove reason
and press ENTER
to clear message
Temp. Sensor int. SHORT!
please remove reason
and press ENTER
to clear message
Temp. Sensor int. OPEN!
please remove reason
and press ENTER
to clear message
check reservoir fluid level.
if reservoir level normal, see Section
3, contact our Sales, Service and
Customer Support.
check reservoir fluid level.
if reservoir level normal, see Section
3, contact our Sales, Service and
Customer Support.
Analog/Digital converter error,
contact our Sales, Service and
Customer Support.
check connections on the back of the
unit
NOTE
before connecting ensure the
power is off.
contact our Sales, Service and
Customer Support.
shorted internal temperature sensor
contact our Sales, Service and
Customer Support.
open internal temperature sensor
contact our Sales, Service and
Customer Support.