Functional Tests

Test Equipment

Test Equipment

The table below summarizes the test equipment needed to perform all of the
functional tests. Alternate equipment model numbers are given in case the
recommended equipment is not available. If neither the recommended nor the
alternative test equipment are available, substitute equipment that meets or
exceeds the critical specifications listed.
Table 1-1
Analyzer
Option
Adapters
503, 508,
3.5 mm (f) to 3.5mm (f)
513, 526
(connector saver for source)
543, 544,
2.4 mm (f) to 2.4mm (f)
550
(connector saver for source)
543, 544,
2.4 mm (f) to 3.5mm (f)
550
All
BNC (f) to SMA (m)
All
Type N (f) to Type N (f)
All
Type N (m) to 3.5 mm (m)
All
Type N (m) to 3.5 mm (f)
All
Type N (f) to 3.5 mm (f)
All
Type N (m) to BNC (f)
Attenuators
All
10 dB Step Attenuator
All
10 dB Fixed Attenuator
(2 required)
Chapter 1
Item
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 50 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 40 MHz
Frequency: 10 MHz to 18 GHz
VSWR: < 1.05:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 1 GHz
VSWR: < 1.05:1
Frequency: 50 MHz
Range: 0 to 70 dB
Frequency: 50 MHz
VSWR: < 1.2:1
Critical Specifications
Al ternate
Recommended
Keysight
Keysight Model
Model
1250-1749
83059B
11900B
11901B
1250-1200
1250-1472
1250-0777
1250-1743
1250-1744
1250-1745
1250-1476
8495A
8496A
8493C
8493A
Option 010
Option 010
or
8493B
Option 010