9
10 Select OK to close the "Analyzer Setup" window.
11 Switch to the Listing window by selecting the Listing tab at the bottom of the main logic analyzer
12 Select the Run Repetitive icon
13 Determine pass or fail (2 of 2 tests) occurrences. See

Other Pods

Repeat the above steps for testing pods 4 to 8. Upon completion, the logic analyzer is completely
tested.
Conclude the State Data Rate Tests
Do the following steps to properly shut down the logic analyzer session after completing the state
mode tests.
1
Keysight U4164A Logic Analyzer Service Guide
In Eye Scan - Open Assign Busses and assign 4 bits to My Bus.
Determine pass or fail (1 of 2 tests) Eye Scan Location. See
window.
End the test.
a From the Main Menu, choose File→Exit. At the dialog "Do you want to save the current
configuration?" select No.
b Disconnect all cables and adapters from the pulse generator.
page
.
page
45.
Testing U4164A Performance
42.
3
47