Function A: Detector Settings

8.2.2 Exciting Current Value (Function A-1)

The exciting current value can be checked/changed by the following procedures.
Be sure to match the exciting current value with the value specified for the
combined detector.
Specifying any other value may cause an error.
Shown below is an example of changing the exciting current value from 0.1900A to 0.2150A.
Switch operation
STEP1
STEP2
STEP3
STEP4
STEP5
Display example
A:DETECTOR
Select "EX CUR" from the setting item
1 EXC CUR
selection menu.
2 SIZE
3 EXC FREQ
4 FLOW DIR
5 EXIT
ESC
EXC CUR
The currently set excitation current value
(0.1900A in this example) appears.
0.1900
Then push
A
* Pushing
ESC
EXC CUR
The switches at the bottom change.
(
0.1900
At the same time, the cursor appears.
(The digit indicated by the cursor is
A
highlighted.)
Then push
EXC CUR
You can continue to change the setting
value. Push
0.1900
the digit you want to change. (You can hold
down the switch longer for continuous
A
operation.)
Then push
EXC CUR
You can continue to change the setting
value. Pushing
0.2900
number of the digit the cursor is positioned.
(You can hold down the switch longer for
A
continuous operation.)
EXC CUR
Repeat this operation to change the value to
0.2150A.
0.2150
When the desired value is obtained, push
A
- 55 -
Description
.
ESC
returns you to the setting
menu.
are shown.)
.
to move the cursor to
.
to set the value temporarily.
6 F 8 A 0 9 9 7