Wiring
4.1 

Prepare transmitter conduit openings

The transmitter enclosure has six conduit openings. Four conduit openings are fitted with
conduit plugs.
Conduit openings accept ½-in. conduit fittings or PG 13.5 cable glands. To keep the case
watertight, block unused openings with NEMA
4.2 
Provide power to the sample conditioning
system
Complete the following steps to power the sample conditioning system.
 WARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electric Code (ANSI/NFPA-70)
and/or any other applicable national or local codes.
Note
Provide a switch or breaker to disconnect the sample conditioning cabinet from the
main power supply. Install the switch or breaker near the unit and identify it as the
disconnecting device for the sample conditioning system.
Procedure
1. Be sure the pump switches on the wiring access panel are in the Off position.
2. Remove the four screws securing the wiring access panel. Pull the panel out of the
3. Insert the power cable through the strain relief connection labeled Power.
4. Wire the power cable to the terminal strip as shown in
Rosemount TCL 56
way to reveal the power terminal strip.
See
Figure
3-2.
 CAUTION
EQUIPMENT DAMAGE
Do not apply 230 Vac power to a 115 Vac TCL (Model option -11). Doing so will
damage the instrument.
4X or IP65 conduit plugs.
®
Figure
4-1.
Wiring
January 2023
17