Data and event logging and retrieval
January 2023
3. Make the appropriate date and time settings and choose which alarm relay
9.3 

Downloading data and events

Complete the following steps to download data and events from your 56 Transmitter to a
USB flash drive.
Procedure
1. To download data or events, move the cursor to the Download tab and press
2. Unscrew the USB port cover in the lower right hand corner of the front panel and
3. Press the appropriate button to download data or events.
Downloaded data and events are stored in a spreadsheet. There is a separate spreadsheet
for every day of data. The finename for downoloaded data is dl mmddyy or dl ddmmyy
depending on the date and time format you select. The filename for downloaded events is
el mmddyy or el ddmmyy.
9.4 

Date and time

To reset the date and time from the main Menu, press the Time and Date button.
Note
Setting the date and time to an earlier value than the one showing will cause data to be
lost from the data/event logger. See
46
activations and deactivations to record as events.
Note
Setting the date or time to an earlier value than the one currently showing will cause
data to be lost form the data/event logger. Download data before resetting time or
date. See
Downloading data and
ENTER/MENU.
insert a USB flash drive in the port.
Downloading may take as long as 20 minutes. During download, the display and
keypad are frozen, but all other transmitter functions continue.
events.
Downloading data and
events.
Reference Manual