Instrument Messages
Introduction
Advisories
Advisory messages tell the front panel user some useful information. (For example,
"File saved successfully" or "Measuring the fundamental".)
Advisory messages appear in the Status Panel at the bottom of the display. The
message remains until you press a key, or another message is displayed in its place.
Advisory messages are not logged in the error queues.
Grayout messages are a special type of Advisory, which appear when you attempt to
access a function that is not available. This could be a grayed out front panel key, or
an inappropriate SCPI command. There are two types of grayout messages: Benign
and Forced.
1. Benign: the requested function is not available because it does not make sense
with the current instrument settings. Changing it does not affect the current
measurement. (For example, setting the number of FFTs/Span when you are not in
the FFT mode.)
A benign grayout gives an Advisory type of message only when the front panel key
is pressed.
The requested function cannot be changed from the front panel, but it can be
changed remotely.
2. Forced: the requested function is not available either because changing it would
cause an invalid measurement, or because of hardware limitations, or because the
selection conflicts with other settings. (For example, selecting the electrical
attenuator when the frequency span includes frequencies above 3.6 GHz.)
A forced grayout function cannot be changed either from the front panel or
remotely. It generates a special type of Advisory message. It also only appears on
the front panel when the key is pressed. Remotely, the message will appear in the
event queue as a warning "-221, Settings conflict; <conflict description>".

Event Message Format

The event messages are listed in numerical order according to their message
number. Advisory messages do not have numbers, and are listed in
alphabetical order.
An explanation is included with each error to further clarify its meaning. Some
errors are specified in industry standards, and may include references to
Section 21.8 of the 1999 SCPI Syntax & Style Standard
(http://www.ivifoundation.org/docs/scpi-99.pdf).
Keysight N9010B EXA Signal Analyzer Service Guide
71