Proceq PM8500 User Manual page 28

Profometer corrosion
Table of Contents

Advertisement

Potential Shift
If Auto Potential Shift is not enabled, the potential readings will be corrected considering this
value.
Auto Save
This feature only applies to the Rod Electrode.
If selected, then measured values of < -50mV will be acquired automatically once the value
has stabilized. If automatic measuring is not selected, then values must be acquired manually.
In both cases, values > -50mV must be acquired by manually.
Auto Save Delay
If Auto Save is selected, this feature allows the user to manually adjust the time of stabilization
between two consecutive measurements that are automatically acquired with the auto save
mode.
Potential Values
It shows and hides the potential values of each cell.
Grid Lines
It shows and hides the grid lines.
Aspect Ratio 1:1
It applies an aspect ratio 1:1 for zooming and un-zooming.
Cell Spacing:
Grid X Spacing It determines the horizontal width of the measured cell (wheel electrode) or the
horizontal distance between measured points (rod electrode).
Grid Y Spacing It determines the vertical distance between two adjacent measuring lines
(wheel electrode) or the vertical distance between measured points (rod electrode). In case of
four-wheel electrode, it should be set equal to the distance between the wheels (from 150 mm
to 250 mm or custom).
Measurement Path
It determines the direction of the measuring sequence when using the rod or the wheel
electrode (Left to right, bottom to top, horizontal zig zag, vertical zig zag).
Fixed Grid
This applies both to the rod electrode and the wheel electrode.
The grid dimension is prefixed but can be defined by the user.
25. September 2023