Proceq PM8500 User Manual page 27

Profometer corrosion
Table of Contents

Advertisement

Measuring Mode
Select between Basic and Advanced.
-
Basic Mode: Directly shows the chipping graphic during the measurement process.
The threshold limits can either be selected manually or using the ASTM standard.
-
Advanced Mode: recommended for expert users. It will display the potential map, and
then will present a statistical analysis of all the potential values measured. Taking
these values into account, the threshold limits can be selected, and therefore, the
chipping graph created.
Threshold standard
If basic mode is selected, this additional feature will appear. Threshold limits can be selected
either manually or using the ASTM standard (passive threshold = -200mV, corrosion threshold
=-350mV).
Probe Type
Select between Rod, One Wheel and Four wheel depending on the electrode connected.
Half-Cell Solution
Select between Cu/CuSO4 (Copper / Copper sulphate), Ag/AgCl (Silver / Silver chloride) or
Hg/Hg2 Cl2 (SCE Calomel).
Unit
Select Metric or Imperial.
Concrete Temperature
Input the temperature of the concrete surface that is being tested.
Auto potential shift
Enabling this feature will correct the potential readings automatically based on the concrete
temperature and the selected electrode.
Disabling this feature will correct the potential readings manually taking into consideration the
potential value included under the "Potential Shift" setting.
25. September 2023
Figure 25 : PM8500 Profometer App: potential shift according to temperature
© Copyright 2023, PROCEQ SA