  • Page 1 Courier 7000 Instruction Manual...
  • Page 3 REVISION HISTORY Date Version Description March 30, 2023 Initial release July 7, 2023 Updated the company address in the back cover and FCC page...
  • Page 5: Table Of Contents

    Courier 7000 EN-1 TABLE OF CONTENTS INTRODUCTION ............................3 1.1. Safety Precautions ...........................3 1.2. Intended Use ............................3 1.3. Overview of Parts and Controls .......................4 1.4. Control Functions .............................4 INSTALLATION ............................6 Packing List ..............................6 Unpacking the Scale ..........................6 Setup ................................7 2.3.1 Power Supply Connection .........................7 2.3.2...
  • Page 6 Table of GEO Code Values ........................49 COMPLIANCE ............................50 APPENDICES ............................52 11.1 MT Standard Continuous Output ......................52 11.2 MT-SICS Commands ..........................54 11.3 OHAUS Commands ..........................55 11.4 OH-Continuous Print ..........................56 11.5 Column Kit Installation........................... 57 LIMITED WARRANTY ............................58...
  • Page 7: Introduction

    Any other type of use and operation beyond the limits of technical specifications, without written consent from OHAUS, is considered as not intended. This instrument complies with current industry standards and the recognized safety regulations;...
  • Page 8: Overview Of Parts And Controls

    Figure 1-1 i- Courier 7000 Item Description Item Description Front Housing RS232 Connector Control Panel USB Type-C Connector Rear Housing Mini DIN Connector Screws (4) Load Cell Connector Battery Cover 1.4. Control Functions Figure 1-2 i- Courier 7000 Control Panel...
  • Page 9 Courier 7000 EN-5 Button Primary Function On/Zero Print Function Tare (Short Press) If the scale is off, If an interface is This key initiates Performs a tare short press to turn enabled, the current the specific operation. on the scale.
  • Page 10: Installation

    Remove the foam blocks. Foam block Take the Courier 7000 scale out of the box and put it on a stable surface. NOTE: For i-C71M150X, please refer to the pictures on the right to carry the Courier 7000 scale.
  • Page 11: Setup

    Courier 7000 EN-7 Setup 2.3.1 Power Supply Connection The Type-C cable and the AC adapter can supply power to the scale. USB Power Use the Type-A end of the USB cable to connect to the Mounting hole Use the other side of the USB cable to connect the Type A terminal through the mounting hole in the bracket.
  • Page 12: Direct Wall Mounting

    2.3.4 2nd Display Installation (available as option) OHAUS offers a remote 2nd Display as an option which can be either mounted to a wall or desk, or mounted on an optional column. Commonly this is used for displaying weight results to a customer in a retail application.
  • Page 13 Column kit All the options can work via the Mini DIN connector on the Courier 7000 scale. Refer to the figure 2-1, you can find the pinouts definition for the Mini DIN. Figure 2-1 Courier 7000 Mini DIN...
  • Page 14: Leveling The Equipment

    EN-10 Courier 7000 Leveling the Equipment To ensure accuracy, proper performance and safety, locate and operate the base on a stable, level surface. Avoid locations with rapid temperature changes or excessive dust, air currents, vibrations, electromagnetic fields or heat. Level the base by adjusting the four leveling feet until the bubble in the level indicator is centered. A wrench may be needed to loosen the locking nut above each leveling foot.
  • Page 15: Operation

    Courier 7000 EN-11 3. OPERATION Turning On/Off the Scale To turn on the scale, you can press the On button. The scale performs a display test, and momentarily displays the software version, the GEO value and then enters the last active weighing mode.
  • Page 16: Totalization

    EN-12 Courier 7000 3.5.2 Totalization This application is used to accumulate the multiple weights manually or automatically. Statistical data (the number of samples, the total weight, the average weight, the minimum weight, the maximum weight and the difference in weight) is stored in memory for review and printing.
  • Page 17: Dynamic Weighing

    Courier 7000 EN-13 Printing Totalization Result and Format tABLE Refer to the steps below, you can enable the Table on ( ) option in the menu. mM.E.N.U Long press the Menu button until you see C.A.L Release the Menu button and .
  • Page 18 EN-14 Courier 7000 Entering the Dynamic Weighing Mode To enter the dynamic weighing mode, you can do the steps that follow. dyNamM Long press the button Function Mode until appears on the display. rEAdY Release the button Function Mode, and appears on the display.
  • Page 19: Weight Alert

    Courier 7000 EN-15 3.5.4 Weight Alert This application is used to judge if the weight of the item is within the target limit. There are 2 limit values for setting, the limit 1 is normally open as default. Setting the Limit Values...
  • Page 20: Menu Settings

    EN-16 Courier 7000 4. MENU SETTINGS The User Menu allows the users to customize the scale settings. Note: Sub-menu for options (USB, Ethernet and Wireless in the table below) will be active when the specific board is installed. Refer to the accessory manuals, you can find the additional setting information.
  • Page 21: Button Navigation

    Courier 7000 EN-17 Some modes/units may not be available in all models. When LEGAL FOR TRADE is turned ON (the security switch S1 is in the locked position), the menu settings will be affected as below: • Calibration function is disabled.
  • Page 22: Initial Calibration

    EN-18 Courier 7000 4.2.1 Initial Calibration When the scale is operated for the first time, a zero and span calibration are recommended to ensure accurate weighing results. Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 4-1.
  • Page 23: Linearity Calibration [Lin]

    Courier 7000 EN-19 NOTE: CAL E • If the scale does not perform the calibration successfully, is displayed for 3 seconds and the previous calibration data is restored. CAL E • After waiting for 40 seconds, if the scale still fails to perform the calibration, is displayed for 3 seconds, and the previous calibration data is restored.
  • Page 24: Calibration Test [C.test]

    EN-20 Courier 7000 4.2.6 Calibration Test C.test Calibration test procedures: mM.E.N.U Long press the Menu button until you see C.A.L Release the button and wait for the display to show . Press the Yes button. C.test Short press the No button several times to navigate until you see .
  • Page 25: Readout Menu

    Courier 7000 EN-21 Capacity Set the capacity of the scale or the first capacity of the scale. 1-999999 GrAd Grad Set the scale readability or the first readability of the scale. 0.0001~100 P.ZerO Power On Zero Zero the scale at Power On.
  • Page 26 EN-22 Courier 7000 StAblE Stability Set the amount reading can vary before the stability symbol turns off. 0.5d = 0.5 of a scale division = 1 scale division = 2 scale divisions = 5 scale divisions [ZErO] Zero Set the percentage of scale capacity that can be zeroed.
  • Page 27: Unit Menu

    Courier 7000 EN-23 Unit Menu U.N.I.t You can activate the desired units with the menu Reset Gram (g) Kilogram (kg) Pound (lb) Ounce (oz) Pound:Ounce (lb:oz) NOTE: Available units vary by model. In addition, due to national laws, the terminal may not include some of •...
  • Page 28: Print Menu

    EN-24 Courier 7000 StOP Stop bit Set the number of stop bits. 1 bIt = 1 stop bits = 2 stop bits [H.SHAKE Handshake Set the flow control method. Hardware handshaking is only available for COM1 menu. NONE = no handshaking...
  • Page 29 Courier 7000 EN-25 ASSIGN Assignment Demand = printing occurs when Print key is pressed. On Stable = printing occurs each time the stability criteria are met On Accept = printing occurs each time the display is within the Weight Alert accept range and stability criteria are met.
  • Page 30 = the Tare weight is printed Header Set the status. = disabled = the Header is printed NOTE: Refer to 11.3 OHAUS Commands, you will know how to enter the header line. Footer Set the status. = disabled = the Footer is printed Mode Set the status.
  • Page 31: Lock Button Configuration

    Therefore, you can see the net weight and the gross weight from the output. NOTE: You can contact an authorized OHAUS dealer to install the ScaleMate software and learn more about the software's other functions. Lock Button Configuration L.o.c.k...
  • Page 32: Legal For Trade

    Refer to the steps below, you can turn the security switch on. 1. Disconnect the power from the terminal. NOTE: Remove the dry batteries of Courier 7000 if used. 2. Remove the housing. 3. Move the security switch (S1) to ON. Refer to the following graphic for the location of S1.
  • Page 33: Sealing

    -- Wire Sealing: replace the countersunk screw by a cross-drilled screw that can be found in the packaging box. Then thread a sealing wire through the screw head and a permanent tab on the bottom housing shown as Figure 5-3. Figure 5-2 Courier 7000 Paper Sealing Item Description...
  • Page 34: Communication

    EN-30 Courier 7000 6. COMMUNICATION The Courier 7000 scale can be set up to be compatible with the certain third party shipping software. UPS WorldShip / FedEx Software Compatibility 6.1.1 UPS WorldShip The OHAUS UPS WorldShip Update Tool allows the UPS World software to recognize the scale. You can download the tool via this website: http://dmx.ohaus.com/WorkArea/showcontent.aspx?id=4295011443...
  • Page 35 Click Apply button to save the selections. Click OK button to close the System Preferences Editor window. Select the Company or Name from the drop-down list. Courier 7000 is ready for UPS weight processing. NOTE: UPS WorldShip software only processes the numeric weight value from the scale -- it does not autosense the weighing unit (kg, lb or oz) nor does it make a conversion from one weighing unit to another.
  • Page 36: Fedex

    EN-32 Courier 7000 6.1.2 FedEx Shipping Software Scale to PC connection Protocol Settings Terminal Settings FedEx USB HID Install the FedEx Ship Manager software on the PC. Start the FedEx Ship Manager software and go to Utilities to select Configure Scale. A window called configure scale appears on the screen.
  • Page 37 Courier 7000 EN-33 Shipping Software Scale to PC connection Protocol Settings Terminal Settings Baud rate: 4800; Parity: Even; RS232 Data Bit: 7; Stop bit: 1 bit FedEx 3835 USB CDC FedEx Ship Manager provides access only to COM ports 1-9 when using RS-232 to USB converter, assign the USB COM port through Device Manager to reflect as COM port 1-9 in advanced settings.
  • Page 38: Scale To Device Port Connections

    If you need to connect the host computer software via the virtual serial port, you may connect the scale to a PC by using the USB Type C cable and selecting the CDC type connection in the Courier 7000 software: Steps Long press the Menu button until mM.E.N.U appears on the display.
  • Page 39: Rs232 Port

    Short press the Yes button to confirm. Courier 7000 RS232 Connection You can connect the optional RS232 cable to the RS232 connector on the Courier 7000 terminal. Refer to figure 6-1, you can find the definition of each pin. RS232 Connector...
  • Page 40: Other Options (2Nd Rs232 Or Ethernet Or Wifi / Bluetooth Or 2Nd Display)

    EN-36 Courier 7000 6.2.3 Other Options (2nd RS232 or Ethernet or WiFi / Bluetooth or 2nd Display) The scale can be set up to print and communicate with the other device via the 2nd RS232 or Ethernet or WiFi / Bluetooth.
  • Page 41: 8213 Interface Protocol

    Courier 7000 EN-37 Request scale to tare Command: T<CR> (54h,0dh) simulate TARE key <LF> <H1> <H2> <CR> <ETX> Change units of measure Command: U<CR> (55h,0dh) simulate UNIT key <LF> <U> <U> <CR> <LF> <H1> <H2> <CR> <ETX> (Ib/kg) simulate UNIT key <LF>...
  • Page 42 EN-38 Courier 7000 in Ib/oz/kg/g <STX> <W> <W> <W> <W> <W> <U> <U> <CR> in Ib/oz/kg/g <STX> <W> <W> <W> <W> <W> <U> <U> <CR> in Ib/oz/kg/g <STX> <sp> <W> <W> <W> <W> <W> <U> <U> <CR> in Ib:oz <STX>...
  • Page 43 Courier 7000 EN-39 Confidence <C> always 0 always 0 always 0 0=RAM ok Bit Definition 1=RAM error 0=Flash ROM ok 1=Flash ROM error always 0 always 0 parity Unrecognized Command Command: others Response <STX> <S> <CR> Definition of the Symbols and the Bits <STX>...
  • Page 44: 3835 Protocol

    EN-40 Courier 7000 6.3.3 3835 Protocol 3835 Protocol Request displayed weight Command: W <CR>(57h, 0dh) over capacity <LF> <U> <U> <CR> <H1> <H2> <ETX> (invalid data) under capacity <LF> <U> <U> <CR> <H1> <H2> <ETX> (-20d) in Ib/oz/kg/g <LF> <p> <W>...
  • Page 45: Nci-Scp01 Protocol

    Courier 7000 EN-41 6.3.4 NCI-SCP01 Protocol Command: W<CR> (57h 0dh), request current reading Response <LF>^^^^^^^^U <CR><LF>H <CR><ETX>---over capacity <LF>_ _ _ _ _ _ _ _U <CR><LF>H <CR><ETX>---under capacity <LF>- - - - - - - - U <CR><LF> H <CR><ETX>---zero-point error...
  • Page 46 EN-42 Courier 7000 Transaction String The length of each item in a transition string. Reading data--6 bytes  Data polarity---1 byte: “-“ for negative, and followed the first digit, “ ” for positive.  Decimal point--1 byte: “.“  Measure unit---1-5 bytes: “Ib”, “kg”, “Ib:oz”, “pcs”, “%”. Units are always lower case, left aligned.
  • Page 47: Maintenance

    7. MAINTENANCE Cleaning For Courier 7000, the housing may be cleaned with a cloth dampened with a mild detergent if necessary. CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING. WARNING: Electric Shock Hazard. Disconnect the equipment from the power supply before cleaning.
  • Page 48: Accessory

    EN-44 Courier 7000 8. ACCESSORY Item Description Picture Platform i-C71 Balltop for R base Platform i-C71 Rollertop for R base Platform i-C71 Balltop for X base Platform i-C71 Balltop for L base Platform i-C71 Rollertop for L base Platform i-C71 Rollertop for X base...
  • Page 49 Courier 7000 EN-45 Column Kit i-C71 CS RS232 Interface Kit Wi-Fi / Bluetooth Kit i-C71 Ethernet Interface i-C71 Auxiliary display i-C71 SF40 Printer Cable RS232 IBM 9P Male-to-Female...
  • Page 50: Technical Data

    EN-46 Courier 7000 9. TECHNICAL DATA Specifications Equipment Ratings: Indoor use only Altitude: 2,000m/6,562ft Operating temperature: -10°C to 40°C (14°F to 104°F) Humidity: Maximum relative humidity 80% for temperatures up to 31°C (87.8°F) decreasing linearly to 50% relative humidity at 40°C (104°F).
  • Page 51 Courier 7000 EN-47 Table 9-1 AM Models Specification Model i-C71M15R i-C71M75R i-C71M50L i-C71M125L i-C71M125X 15 kg x 0.005 kg 75 kg x 0.02 kg 50 kg x 0.01 kg 125 kg x 0.02 kg 125 kg x 0.02 kg Capacity ×...
  • Page 52: Drawings And Dimensions

    EN-48 Courier 7000 Table 9-2 Other Models Specification Model i-C71M15R i-C71M60R i-C71M60L i-C71M150L i-C71M150X Capacity × 15 kg x 0.002 kg 60 kg x 0.01 kg 60 kg x 0.01 kg 150 kg x 0.02 kg 150 kg x 0.02 kg...
  • Page 53: Table Of Geo Code Values

    Courier 7000 EN-49 Table of GEO Code Values Table 9-4 GEO Codes Elevation in meters 1300 1625 1950 2275 2600 2925 3250 1300 1625 1950 2275 2600 2925 3250 3575 Elevation in feet 1060 2130 3200 4260 5330 6400 7460...
  • Page 54: Compliance

    This product complies with the applicable harmonized standards of EU Directives 2011/65/EU (RoHS), 2014/30/EU (EMC), 2014/35/EU (LVD) and 2014/31/EU (NAWI). The EU Declaration of Conformity is available online at www.ohaus.com/ce. This product complies with the applicable statutory standards of the Restriction of the Use of...
  • Page 55 Unintentional Radiator per 47CFR Part B Trade Name: OHAUS CORPORATION Model or Family identification: i-C71 Party issuing Supplier’s Declaration of Conformity: Ohaus Instruments (Changzhou) Co., Ltd. Building C, No. 6 Zhengqiang Road, Xuejia Town, Xinbei District, Changzhou Jiangsu 213022 China Phone: +86 519 85287270 Responsible Party –...
  • Page 56: Appendices

    Non-significant weight data and tare data digits are transmitted as spaces. The continuous output mode provides compatibility with OHAUS products that require real-time weight data. the standard continuous output. Table 11-1 shows the format for the standard continuous output.
  • Page 57 Courier 7000 EN-53 Table 11-4: Status Byte C Bit Definitions Bits 2, 1, and 0 Weight Description lb or kg, selected by Status Byte B, bit 4 grams (g) metric tons (t) ounces (oz) not used not used tons (ton)
  • Page 58: Mt-Sics Commands

    EN-54 Courier 7000 11.2 MT-SICS Commands Command Function LEVEL 0 Reset the scale Inquiry of SICS level and SICS versions Inquiry of scale data Inquiry of scale software version Inquiry of serial number Send stable weight value Send weight value immediately...
  • Page 59: Ohaus Commands

    Courier 7000 EN-55 11.3 OHAUS Commands Commands listed in the following table will be acknowledged by the scale. The scale will return “ES” for invalid commands. Please add \r\n after each command to send(\r refers to \return; \n refers to \newline)
  • Page 60: Oh-Continuous Print

    EN-56 Courier 7000 11.4 OH-Continuous Print Format 1: For the printout result including interval and continuous printing mode of Weight Alert application  Weight Unit Stability Field Space Space (Right aligned) (Right aligned) Length T/N/G/PT Application Status Field Space Space Term.
  • Page 61: Column Kit Installation

    Courier 7000 EN-57 11.5 Column Kit Installation...
  • Page 62: Limited Warranty

    OHAUS products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period. During the warranty period OHAUS will repair, or, at its option, replace any component(s) that proves to be defective at no charge, provided that the product is returned, freight prepaid, to OHAUS.
  • Page 63 With offices worldwide / Con oficinas en todo el mundo / Avec des bureaux partoutdans le monde / MitBürosweltweit / Con uffici in tuttoilmondo www.ohaus.com *30809929* P/N 30809929 B © 2023 Ohaus Corporation, all rights reserved / todos los derechosreservados / tousdroitsréservés / alle Rechte vorbehalten / tutti i dirittiriservati...

This manual is also suitable for:

I-c71m15rI-c71m75rI-c71m50lI-c71m125lI-c71m125x