Linearity Calibration [Lin]; Geo Code Adjustment [Geo] - OHAUS Courier 7000 Instruction Manual

Hide thumbs Also See for Courier 7000:
Table of Contents

Advertisement

Courier 7000
NOTE:
• If the scale does not perform the calibration successfully,
previous calibration data is restored.
• After waiting for 40 seconds, if the scale still fails to perform the calibration,
seconds, and the previous calibration data is restored.
4.2.4 Linearity Calibration
The scale performs linearity calibration with 3 calibration points. The full calibration point is established with a
weight on the platform. The mid calibration point is established with a weight equal to half of the full calibration
weight on the platform. The zero calibration point is established with no weight on the platform. The full
calibration and mid calibration points can be altered by users during the calibration procedure.
Calibration procedures:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
3.
Short press the No button several times to navigate until you see
0
4.
kg and the calibration unit are flashing on the display. With no weight on the platform, press the Yes
button to establish the zero point. The display shows
5.
Then the first calibration point and the unit are flashing on the display based on the scale capacity and unit
you set in the setup menu. (For example,
skip to step 7.
6.
To change the calibration point, you can do the steps that follow.
a)
Short press the No button several times until the desired number appears.
NOTE: press the Back button can decrease the digit.
b)
Short press the Yes button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the Yes button to accept calibration point. It is flashing on the display.
7.
Place a calibration mass of the specified weight on the platform and press the Yes button. The display
--C--
shows
.
8.
The second calibration point and the unit are flashing on the display based on the capacity and unit you
set in the setup menu. (For example,
NOTE: After waiting for 40s, if the scale still fails to perform the calibration,
seconds and the previous calibration data is restored.
9.
Repeat step 6 and 7.
10. The display shows
11. After that the display shows
4.2.5 GEO Code Adjustment
Geographical Adjustment Factor (GEO) code is used to adjust the calibration based on the current location.
Settings from 0 to 31 are available with 12 being the default.
Refer to 9.3 Table of GEO Code Values to determine the GEO factor that corresponds to the terminal's
location.
To set the GEO factor, you can do the steps that follow.
Long press the Menu button until you see
1.
2.
Release the button and wait for the display to show
3.
Short press the No button several times to navigate until you see
4.
The GEO point is flashing on the display (For example,
5.
Short press the No button several times until the desired GEO number appears. Press the Yes button to
accept.
NOTE: Press the Back button can decrease the digit.
6.
Then the display shows
[
]
LIN
015.000
030.000
--C--
-DonE-
, and then
GEO
. Press the Exit button to exit.
[GEO]
C.test
. Press the Exit button to exit.
CAL E
mM.E.N.U
.
C.A.L
. Press the Yes button.
--C--
.
kg). If you do not need to change the calibration point,
kg)
when the Linearity calibration is finished.
mM.E.N.U
.
C.A.L
. Press the Yes button.
12
).
is displayed for 3 seconds and the
CAL E
is displayed for 3
LIN.
Press the Yes button.
CAL E
is displayed for 3
GEO.
Press the Yes button.
EN-19