Initial Calibration; Zero Calibration [Zero]; Span Calibration [Span] - OHAUS Courier 7000 Instruction Manual

Hide thumbs Also See for Courier 7000:
Table of Contents

Advertisement

EN-18

4.2.1 Initial Calibration

When the scale is operated for the first time, a zero and span calibration are recommended to ensure accurate
weighing results.
Before performing the calibration, be sure to have the appropriate calibration weights as listed in table 4-1.
Ensure that the LFT switch/calibration lock is set to the unlocked position.
You can also adjust the GEO setting according to your location
the GEO Code Adjustment section below.
Max Capacity (kg)
15kg
60kg
150kg
NOTE: The calibration unit is consistent with the capacity unit.

4.2.2 Zero Calibration [ZErO]

The scale performs zero calibration with one calibration point. The zero-calibration point is established with no
weight on the scale. You can use this calibration method to adjust for a different static load without affecting the
span or linearity calibration.
Calibration procedures:
Long press the Menu button until you see
1.
2.
Release the button and wait for the display to show
3.
The display shows
0
kg and the calibration unit are flashing on the display. With no weight on the platform, press the Yes
4.
button to establish the zero point.
5.
The display shows
NOTE: If zero calibration is failed or if after 40 seconds the calibration is still not successful,
displayed for 3 seconds and the previous calibration data will be restored.
6.
Then the display shows

4.2.3 Span Calibration [SpaN]

The scale performs span calibration with one calibration point. It is established with a calibration mass placed
on the scale.
NOTE: Span calibration should be performed after zero calibration.
Calibration procedures:
1.
Long press the Menu button until you see
2.
Release the button and wait for the display to show
Short press the No button to navigate until you see
3.
4.
The calibration point and the unit are flashing on the display based on the scale capacity and unit set in the
030.000
menu. (e.g.
5.
To change the calibration point, you can do the steps that follow.
a)
Short press the No button several times until the desired number appears.
NOTE: press the Back button can decrease the digit.
b)
Short press the Yes button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
Press the Yes button to accept calibration point. It is flashing on the display.
d)
6.
Place a calibration mass of the specified weight on the platform and press the Yes button.
7.
The display shows
8.
Then the display shows
Table 4-1 Suggested Span Calibration Mass (Sold Separately)
Mass* (kg)
15kg
60kg
150kg
ZErO.
Press the Yes button.
--C--
-DonE-
, and then
SPAN
. Press the Exit button to exit.
kg) If you do not need to change the calibration point, skip to step 6.
--C--
-DonE-
, and then
LIN
. Press the Exit button to exit.
.
For detailed GEO information, please check
Max Capacity (lb)
30lb
100lb
150lb
250lb
mM.E.N.U
.
C.A.L
. Press the Yes button.
when the Zero calibration is finished.
mM.E.N.U
.
C.A.L
. Press the Yes button.
SPAN.
Press the Yes button.
when the calibration is finished.
Courier 7000
Mass* (lb)
30lb
100lb
150lb
250lb