StAblE
Stability
Set the amount reading can vary before the stability symbol turns off.
0.5d
= 0.5 of a scale division
1d
= 1 scale division
2d
= 2 scale divisions
5d
= 5 scale divisions
[ZErO]
Zero
Set the percentage of scale capacity that can be zeroed.
2%
= zero range is within +/-2%
100%
= zero range is within +/-100%
[
]
FILtEr
Filter
Set the amount of signal filtering.
LOwW
= faster stabilization time with less stability.
mMEd
= normal stabilization time with normal stability.
HIGH
= slower stabilization time with more stability.
[
]
AZt
AZT
Set the automatic zero tracking feature.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 of a divisions per second has been
exceeded.
1d
= the display will maintain zero until a change of 1 division per second has been exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been exceeded.
[
B.LIGHt
Backlight
Set the display backlight feature.
OFF
= backlight is disabled.
ON
= backlight is enabled and always on.
AUtO
= backlight is disabled after 20 seconds of no activity.
[
SCreeN
Screen Saver
Set whether the screen saver is enabled after the selected time period.
OFF
= screen saver is disabled
1
= the screen saver is enabled after 1 minute of no activity.
2
= the screen saver is enabled after 2 minutes of no activity.
5
= the screen saver is enabled after 5 minutes of no activity.
[
]
A.OFF
Auto Off
Set whether the display enters sleep mode after the selected time period.
OFF
= disabled
1
= the display enters sleep mode after 1 minute of no activity.
5
= the display enters sleep mode after 5 minutes of no activity.
10
= the display enters sleep mode after 10 minutes of no activity.
P.SAUEr
PWM.SAVE
[
Set whether to enable power saving mode after the scale enters standby mode. If it is enabled, it will take
about 3 seconds before the terminal returns to weighing when weight is changed on the platform or you
press the terminal's buttons.
ON
= power saving is enabled.
OFF
= power saving is disabled.
[
ENd
End Readout
Advance to the next menu or return to the top of the current menu.
]
]
]
]
]