OHAUS Courier 7000 Instruction Manual page 18

Hide thumbs Also See for Courier 7000:
Table of Contents

Advertisement

EN-14
Entering the Dynamic Weighing Mode
To enter the dynamic weighing mode, you can do the steps that follow.
1.
Long press the button Function Mode until
2.
Release the button Function Mode, and
Starting Dynamic Weighing
1.
Put the load on the scale.
• If the operation type is manual, you must press the Function button to start the averaging process.
• If the operation type is semi-automatic/automatic, the scale performs the averaging process
automatically when the load weight is bigger than the start weight.
NOTE:
• You can set the value of the start weight via this path:
correct range of the start weight value is 5d - capacity value.
• During the averaging period, the countdown timer decreases in one second increments. If the set
average time is 0s, the countdown timer is not displayed.
• The values on the display are averaged and held on the display when the countdown is completed.
2.
Read the values on the display if the tilde (dynamic) symbol blinks on the display.
3.
Reset the countdown timer:
• If the operation type is manual/semi-automatic, you can press the Function button to reset the
countdown timer when the countdown is running. The display shows
• If the operation type is automatic, you can remove the load from the scale. And the average weight will
still be displayed until the duration time is over. Then the display shows
NOTE:
rEADY
• The
display must be at zero gross or net value to reset the countdown timer.
Application Settings
The Dynamic Weighing Configurations are defined below (Defaults in Bold).
Item
Dynamic Mode
dyNAmM
Dynamic Operation
d.tyPE
Type (
)
Start Weight
D.StArt
(
)
Duration Time
d.tImME
(
)
Average Time
A.tImME
(
)
NOTE: The Dynamic mode cannot be disabled if the scale is in the Dynamic mode currently.
rEAdY
Available Settings
OFF
ON
Off (
) / on (
)
mMAN
Manul (
)/
semMI
Semi-auto (
)/
AUtO
Auto (
)
5d ~ Capacity Weight
1 ~ 10s
0 ~ 30s
dyNamM
appears on the display.
appears on the display.
mM.E.N.U
>
Comments
Count down:
There is a countdown time.
Manual:
The averaging process is started and reset manually.
Semi-auto:
The averaging process is started automatically and reset
manually.
Auto:
The averaging process is started and reset automatically.
Dynamic weighing will start when the load is bigger than the
start weight (for Semi-auto and Auto mode).
It is the time for the display to remain the dynamic weighing
result after the load is removed.
Time in Seconds.
If the average time is 0, the first stable weight (more than or
equal to 5d) will be the result.
mMOde
dyNamM
D.StArt.
>
>
rEAdY
, and start to re-count.
rEAdY
, and start to re-count.
Courier 7000
And the

Advertisement

Table of Contents
loading

This manual is also suitable for:

I-c71m15rI-c71m75rI-c71m50lI-c71m125lI-c71m125x