Source Pressure Module - OERTLI OS 4 Service Manual

Medical equipment
Table of Contents

Advertisement

11.1.24 Source pressure module

VX210152
Quantity
Description
1
Source pressure module
2021-03-23 / Rev. 10
VV016042
STEP 1:
1.
Remove the black ring that holds
the compressor in place. See step 2
of 11.1.7.
2.
Disconnect 2 tubes from the com-
pressor.
3.
Disconnect all cables between the
compressor and the supply PCB.
STEP 2:
4.
Remove the 2 screws that connect
the compressor with the spring sus-
pension.
5.
Take the compressor unit out of the
unit in a sideways direction.
STEP 3:
6.
Remove the tube connectors from
the old compressor module and fix
them to the new module.
94

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vv016042Vc860100Vc860200Vc860300

Table of Contents