Laser Servicing Software; Introduction - OERTLI OS 4 Service Manual

Medical equipment
Table of Contents

Advertisement

9.5

Laser servicing software

The servicing software as well as the access key to the software will be provided after the successful participation of a laser
service training carried out by Oertli.
9.5.1

Introduction

The laser servicing software enables the service engineer to modify the control parameters of the laser and to read out the
history of the device.
The laser servicing software runs on a PC connected to the laser main unit via a USB cable. Every control parameter can be
read out and adjusted by using the servicing software. The only method possible for laser parameter adjustment is the ser-
vicing software. Moreover, history data can only be monitored by means of the servicing software.
DANGER!
Treatment of patient with servicing software. Direct or scattered laser radiation
Eye or skin damage
► Do not use the laser for clinical treatments (treatment of patients) while a PC or the servicing software is
connected to the laser module. If the laser path is not fully shrouded, suitable laser safety glasses must
be worn by all persons in the room.
2021-03-23 / Rev. 10
STEP 2: Dry cleaning
(only if connector end face is not clean)
4.
5.
6.
STEP 3: Wet/dry cleaning
(If the fibre endface is not clean after multiple dry
cleaning procedures, try wet/dry cleaning)
7.
8.
9.
10.
VV016042
Slide back protective cover and advance
cloth to expose a new, clean surface.
Hold the fibre tip lightly against the
cleaning cloth and wipe it using a figure
8 motion. Do not clean over the same
surface more than once.
Inspect the fibre with the inspection
scope. If not clean, repeat dry cleaning
on a new cleaning cloth.
Fold a lint-free wipe.
Moisten one section of the lint-free wipe
with a drop of fibre connector cleaner. Be
sure that a portion of the wipe remains
dry.
Hold the fibre tip lightly against the mois-
tened area of the lint-free wipe and move
it to the dry section of the wipe.
Discard the lint-free wipe and inspect the
fibre endface with the inspection scope.
Repeat the process, if necessary. If the
test fibre is not clean after multiply reple-
tion of the wet/dry cleaning process, it
must be replaced. It is recommended to
keep a replacement fibre.
30