Error Messages - OERTLI OS 4 Service Manual

Medical equipment
Table of Contents

Advertisement

14.2

Error Messages

Error number
Meaning
Supply error
ERROR 1
+12V, +24V, +5V, -5V, +3.3V not within tol-
erance
ERROR 2
Adjust DIA
Pedal not calibrated or incorrectly cali-
brated
ERROR 3
Adjust pedal
ERROR 4
Remote graphic display
Display error
ERROR 5
(RESET key is blinking)
ERROR 7
CORE software
COM error
ERROR 8
(RESET key is blinking)
ERROR 10
Peristaltic pump failure
ERROR 11
HF output stage
ERROR 12
Irrigation pole motor
ERROR 15
Force sensor
ERROR 16
Adjust Venturi sensor
ERROR 17
Adjust force sensor
ERROR 18
Check Venturi pump system
ERROR 20
Extraction deviation
ERROR 21
Adjust Visco sensor
ERROR 22
Adjust air sensor
ERROR 23
VIT PN differential sensor
ERROR 24
LUM1 module
ERROR 25
LUM2 module
ERROR 26
Venturi System
ERROR 31
Liquid level sensor
ERROR 32
Irr valve
ERROR 33
Air/gfi port
ERROR 38
Laser module
ERROR 39
Wireless module
2021-03-23 / Rev. 10
Solution
Replace system unit.
Replace the system unit and notify the manufacturer.
The pedal has to be readjusted as described in the
Section "Service mode operation / Pedal calibration".
- Check the cable to the remote control panel.
- Replace the remote control panel.
- Check the cable from the carrier board PCB to the
control unit.
- Replace the control panel.
Restart device, send Log-Files to manufacturer
- Restart device, send Log-Files to manufacturer
- Check the cable from the carrier board PCB to the
core PCB
- Replace the carrier board PCB
Check the encoder cable and the motor cable to the
peristaltic pump motor.
Replace the system unit and notify the manufacturer.
- Check if irrigation pole is blocked or overloaded
- Check the cables of the drive for infusion pole
- Replace the drive for infusion pole
- Adjust force sensor
- Ascertain whether the force sensor is plugged in or
whether there is a severed cable
- Replace fluidic unit.
Adjust Venturi sensor or replace fluidic PCB.
Adjust force sensor or replace fluidic unit.
- Adjust Venturi and force sensor
- Replace fluidic unit
- Replace compressor unit
- Replace visco module
Adjust injection and extraction sensor.
Adjust AIR sensor and pressport AIR sensor.
Replace VIT-PN module
- Wait for a short time while light module can cool
down.
- Replace LEDplus module, light LEDplus cannot be
operated.
- Wait for a short time while light module can cool
down.
- Replace LED module, Light LED cannot be operated.
- Adjust Venturi Sensor
- Replace Venturi unit
Notify the manufacturer
- Replace irrigation valve
- Replace fluidic unit
- Adjust Control AIR sensor and pressport AIR sensor.
- Replace AIR/GFI module
- Switch off device, disconnect it from the mains sup-
ply, connect it to the mains supply, switch device on
- Notify manufacturer
- Restart device
- Replace wireless PCB
VV016042