7.6

Check - Logging

8

OS 4 service tools

8.1

Conversion of the surgeon memory

The following chapter describes the procedure in the case of an external conversion of a surgeon memory. The following
equipment is required:
Laptop with an USB interface.
USB stick with a database exported from a device with SW version 1.3.0 or higher
Software for converting the surgeon database (OS4_Database_Conversion.exe)
The software for converting the database does not have to be installed onto the laptop. Simply copying the software onto the
laptop or executing it directly from a USB stick is enough.
Convert the surgeon database before performing the software update. Otherwise, in the case of a faulty data-
base (especially as regards pedal assignments), corrections cannot be made anymore!
The converted database must be located on a USB stick in the folder „dbexport" so that it can be imported
into the OS 4. This folder must be at the highest level within the data system of the USB stick. This is set as a
default in the case of a database export from an OS 4 device onto a USB stick as well as after a database
conversion.
The latest version of the software for converting the surgeon database (version 1.3.4 or higher; OS 4_Data-
base_Conversion.exe) can be used for conversion of the surgeon database from software 1.3.0 to software
1.4.0 (8.1.3), software 1.4.0 to software 1.5.x (8.1.2) or software 1.5.x to software 2.0.x (8.1.1).
2021-03-23 / Rev. 10
Export all log files to USB:
VV016042
Plug in USB flash drive
Push the button <Export all log files to
USB>
Unplug USB flash drive
Only use an Oertli service USB
flash drive in conjunction with
the OS 4.