Procedure For Converting A Database From A Device With Sw Version 1.3.0 - OERTLI OS 4 Service Manual

Medical equipment
Table of Contents

Advertisement

8.1.3

Procedure for converting a database from a device with SW version 1.3.0

The conversion tool requires that following a software update to SW 1.3.0 all pedal assignments must be re-
selected for the surgeon memories that have been used. If a database is to be converted that has not previ-
ously undergone this process, a single option is offered to set back to default all pedal assignments of a sur-
geon, of which the pedal assignments have not been reselected.
2021-03-23 / Rev. 10
7.
8.
9.
1.
2.
3.
4.
5.
6.
VV016042
The converted database has been stored
at the place of origin (normally on the
USB stick). The original database has
been stored as a backup copy with the
ending of the corresponding scheme ver-
sion.
The software of the device can now be
updated to software package 1.5.x
(12).
Following a successful software update,
the converted database has to be loaded
into the device according to 7.2.
Export surgeon database from device
with SW 1.3.0 to USB according to
7.2.
Insert the USB stick containing the
exported surgeon database into the
USB port of your laptop.
Start the software for the conversion
of the surgeon database.
(OS 4_Database_Conversion.exe)
Click on the „Browse db-files" button
and select exported database.
After successful loading of the file
(Status „File loaded"), start the con-
version process by clicking on „Start
conversion".
If none of the laser pedal assign-
ments have been correctly set, a
popup window will open up, in which
laser pedal assignments are as-
sessed.
Valid laser pedal assignments:
For these surgeons, the laser
pedal assignments (sections
and buttons) have been cor-
rectly set.
Invalid laser pedal assignments:
For these surgeons, the laser
pedal assignments (sections or
buttons) have not been correctly
set.
There are two possibilities of how to
proceed:
Click on the button "Replace in-
valid laser assignments by de-
fault": Only invalid laser pedal
assignments will be replaced by
the corresponding default set-
tings.
20