13.6.2

Vertical pedal sensor

VX310082 Pedal sensor vertical
Quantity
Description
1
Pedal sensor vertical
2
M3x12 oval head screw, TORX, ecosyn-fix
1
Cable tie
After the exchange of a pedal sensor, the pedal has to be sealed. For this purpose, additionally order
VX102479 Sealing set (see chapter 13.2)!
2021-03-23 / Rev. 10
STEP 1:
1.
2.
3.
4.
STEP 2:
5.
6.
7.
VV016042
Open pedal housing according to the
procedure described in 13.1.
Remove battery pack according to
13.3.
Disconnect cable from PCB.
Undo the grub screw.
Push out the wave vertical
deflection component from the base
body.
Release spring.
Place rocker in such a way that
access is given to both sensors.
Do not lose the washers when
pushing out the wave vertical de-
flection and reassemble them
after repair.
Lubricate the wave vertical deflec-
tion before reassembling the pe-
dal.