7.4
Sensor calibration
7.5
Pedal calibration
2021-03-23 / Rev. 10
Calibration sensors
Store actual offset values of the sensors
displayed by touching <Calibrate>.
Calibration pedal
Pedal Adjustment:
Pedal Pairing:
Establish radio communication:
Turn off radio communication:
VV016042
DANGER!
Incorrect calibration
Risk of eye damage to pa-
tient
► No instruments shall be
connected to the device
while calibrating sensors.
Before adjusting and pairing the
pedal make sure that it is con-
nected with cable to the OS 4.
Turn off the device, wait for 20 sec. at
the minimum
Connect pedal with cable to the OS 4
Turn on the device
Call up the service mode (see 7.1)
The pedal adjustment has to be
carried out before the pedal
pairing is started.
Make sure that the foot pedal is
placed on the ground.
Push <Start> to begin with the cali-
bration routine.
Move to pedal position shown on the
sequence and <Confirm>.
In the menu <Calibration> select <Pe-
dal>
Push the button <Start Pairing>
As soon as the pairing is finished, the
message "Success" appears.
Connect pedal with cable to the OS 4
Push the button <Enable Wireless>
Connect pedal with cable to the OS 4
Push the button <Disable Wireless>
17