Messages, warnings and error messages / fault correction

14.1

Information and warnings

Message
HF test successful
Surgeon name must be unique
Surgeon name is too long
No pedal
Check phaco handpiece
Phaco handpiece defective
Repeat phaco test
Check phaco tip
Check air pressure
Visco pressure deviation
Temperature to high
No cassette
Laser key switch off
Laser emergency switch off
Laser: no fibre detection - fully connect fibre. Insert fibre into laser port.
Laser door interlock not closed
Remind: Laser user protection
Laser power too low
No serial number on print
Serial number invalid
Pedal not laser conform
Cassette full
HF tip defective
Laser in saturation
Laser: output power too low
Laser beeper malfunction
Laser microphone malfunction
Laser over temperature
Laser: no auto key detected
Laser: foot switch malfunction
Laser: no foot switch detected
Laser: Active user protection filter detected
although it is not configured. Active user
protection filter is used.
Laser: active user protection filter is plugged
but not in place.
Laser: no active user protection filter de-
tected – connect active user protection filter
2021-03-23 / Rev. 10
Solution
-
Change surgeon name.
Choose a surgeon name with fewer letters.
Connect a pedal or switch pedal on.
Use another handpiece.
Plug in another phaco handpiece or a handpiece.
Repeat phaco test.
- Check whether tip is broken.
- Tighten tip correctly.
- Check whether air pressure is on.
- Plug in air hose.
Check Visco setup.
Wait for a short time while machine can cool down.
Insert a cassette.
Turn the key switch for the laser on.
Release emergency switch.
Insert DRS dongle.
Check whether laser protection filter in microscope is in place and all
persons in the room wear their laser protection glasses.
- Set laser power value.
Call service.
Call service.
Connect an OS 4 pedal.
Empty cassette into a drainage bag or change cassette.
- Change HF tip or HF handpiece.
- Decrease laser power and/or pulse duration.
- Increase interval between two pulses.
- Check fibre and recalibrate laser.
- Call for laser second level support.
Call for laser second level support.
Call for laser second level support.
- Wait until system has cooled down.
- Reduce laser power or increase pulse interval.
- Auto key at the laser module (connector at front of laser module) is
missing. Call service.
Call for laser second level support.
- Connect OS 4 foot pedal and confirm.
- Switch pedal on and confirm.
If active user protection filter is installed permanently, change the config-
uration for the user protection filter in the service mode to "active".
Install the active UPF in the microscope according to the manufacturer's
instructions.
Connect active user protection filter.
If you wish to permanently use a passive user protection filter, install a
passive user protection filter and change the configuration for the user
protection filter in the service mode to "passive".