11.1.6

Front enclosure

2021-03-23 / Rev. 10
VV016042
STEP 1:
1.
Remove upper enclosure and lower
enclosure according to section
11.1.5 steps 1 and 2.
2.
Loosen front enclosure by removing
2 screws.
STEP 2:
3.
Remove the two air lines (green (on
the left side viewed from the back)
and the black) from the valves.
4.
Unplug laser detect JP17 cable from
CORE-Print.
STEP 3:
5.
Pull lower part of the enclosure to-
wards the front, until the light mod-
ules are no longer positioned in the
openings.
6.
Pull front enclosure downwards so
that it comes free from the slot.
Please take great care of the key
switch on the left to avoid break-
age.
7.
Make sure not to tear any cables and
tubes that are still connected to the
part!