Endo Laser System Calibration; Laser Safety - OERTLI OS 4 Service Manual

Medical equipment
Table of Contents

Advertisement

Could not open port
Device not answering
Writing serial number: FAILED
Writing data: FAILED
Checking data: FAILED
Calibrating offset: FAILED
Calibrating gain: FAILED
Spare part not defined
9

Endo laser system calibration

9.1

Laser safety

DANGER!
Direct or scattered laser radiation
Eye or skin damage
► All servicing, adjustments and settings must be carried out by service technicians authorized and trained by
Oertli.
► Use periodically calibrated measurement devices only.
DANGER!
Direct or scattered laser radiation
Eye or skin damage
► Avoid eye or skin exposure to direct or scattered laser radiation.
► Do not turn the laser beam directly towards persons.
► If the laser path within the laser is no longer fully shrouded, suitable laser safety glasses must be worn by
all persons in the room.
► In the case of damage to the laser safety glasses or a change in colouration, discontinue using them. Re-
place them immediately.
NOTE!
Improper repairs
Severe damage to laser module
► Never open the laser module by yourself. Always send the module back to the manufacturer.
2021-03-23 / Rev. 10
Serial interface could not be started.
There is no response from the device.
Error occurred when writing the serial
number into the device.
Error occurred when writing the ad-
justment data into the device.
Faulty calibration data on the device.
Only concerns the fluidics spare part:
Offset could not be calibrated.
Either error in communication or cali-
bration value outside tolerance range.
Only concerns the fluidics spare part:
Gain could not be calibrated. Either
error in communication or calibration
value outside tolerance range.
For this spare part, there is no coun-
terpart in the tool.
VV016042
Select correct COM port (normally "USB
Serial Port")
If correct COM port does not occur in the
list, close tool, check USB interface cable
and restart tool.
Connect the device correctly with the USB
interface cable and switch the device on.
Try to load data onto the device again. In
the case of repeated failure, there may be
a defect on the CORE print. Contact Sup-
port at Oertli Instrumente AG.
Connect required equipment as described
in the tool and start tool again.
Connect required equipment as described
in the tool and start tool again.
Use correct version of the tool.
28