8.1.1

Procedure for converting a database from a device with SW version 1.5.x

8.1.2

Procedure for converting a database from a device with SW version 1.4.0

2021-03-23 / Rev. 10
1.
2.
3.
4.
5.
6.
7.
8.
9.
1.
2.
3.
4.
5.
6.
VV016042
Export complete surgeon database from
device with SW 1.5.x to USB according
to 7.2.
Insert the USB stick containing the ex-
ported surgeon database into the USB
port of your laptop.
Start the software for the conversion of
the surgeon database.
(OS 4_Database_Conversion.exe)
Click on the „Browse db-files" button and
select exported database.
After successful loading of the file (Sta-
tus „File loaded"), start the conversion
process by clicking on „Start conversion".
Successful conversion is indicated by the
display of the status message „Database
export completed" and is confirmed by a
green background of the status field.
The converted database has been stored
at the place of origin (normally on the
USB stick). The original database has
been stored as a backup copy with the
ending of the corresponding scheme ver-
sion.
The software of the device can now be
updated to software package 2.0.x
(12).
Following a successful software update,
the converted database has to be loaded
into the device according to 7.2.
Export surgeon database from device
with SW 1.4.0 to USB according to 7.2.
Insert the USB stick containing the ex-
ported surgeon database into the USB
port of your laptop.
Start the software for the conversion of
the surgeon database.
(OS 4_Database_Conversion.exe)
Click on the „Browse db-files" button and
select exported database.
After successful loading of the file (Sta-
tus „File loaded"), start the conversion
process by clicking on „Start conversion".
Successful conversion is indicated by the
display of the status message „Database
export completed" and is confirmed by a
green background of the status field.
19