8.0

Maintenance

8.1

Cleaning

Caution...
Before cleaning the calibrator, you must switch it off, allow
it to cool down and remove all cables.
Users should/must carry out the following cleaning procedures as and
when required:
The exterior of the instrument - Clean using water and a soft
cloth.
The cloth should be wrung out hard to avoid any water penetrating
the calibrator and causing damage.
When heavily soiled the exterior of the instrument and the
keyboard can be wiped clean using isopropyl alcohol.
The insertion tube - Must always be clean and should be
regularly wiped using a soft, dry cloth.
You must ensure there are no textile fibres on the insertion tube
when it is inserted in the well. The fibres may adhere to the well
and damage it.
The well - Must always be clean. Dust and textile fibres should be
removed from the well using e.g. compressed air.
Warning
REMEMBER! Wear goggles when using compressed
air!