Fiber Recoaters with Linear Proof Testers
Chapter 7: Handset Controller Reference
7.4.1. Purge (PTR406 Only)
Allows the user to initiate a Purge sequence. Purging is a method of
removing air bubbles or voids from the inject mechanism. The system
should be purged whenever a new bottle of material is installed.
Purge Screen
7.4.2. Fill (PTR406 Only)
Allows the user to perform a fill operation. If an Inject sequence is initiated and
the pump syringe has insufficient material to perform the inject operation, the
machine will automatically fill the syringe from the material bottle. In many
situations, the user may prefer to initiate the fill sequence while working away
from the machine. Note that the machine cannot be used for any operation
while performing a fill operation.
Fill Screen
Page 22