Test Card; Preparation - Leica GLOW800 User Manual

Hide thumbs Also See for GLOW800:
Table of Contents

Advertisement

Check and adjust the illumination and functions
9.2

Test card

To check and test the GLOW800 function, to verify the proper
adjustment of the white light and fluorescence image and to verify
the illumination level the GLOW800 test card has to be used.
WARNING
Risk of infection due to unsterile GLOW800 test card
Do not use the GLOW800 test card in the sterile field.
X
Use only in non-sterile environment.
X
Check the microscope illumination in non-sterile
X
environment only.
Take care to ensure the precise parfocal setting of the
X
ARveo 8. Follow the instructions on parfocal setup.
Please prepare the test as follows:
2
Leica GLOW800
10 746 513-01
0
!
4
2
1 Stepwise decreasing NIR Intensity fields 0=bright to 9=dark
2 Low intensity NIR fluorescence area including 4 different white
light color areas
3 Round spot high intensity NIR fluorescence signal
4 Hole to fix the card
26
1
2
1
2
3
4
5
6
7
8
9
Please read the Leica
GLOW800 user manual!
2
GLOW800 / Ref. 10 748 652 / Version 01
5 Color samples for GLOW800 mode and white light color
balancing
6 High intensity fluorescence samples
7 Production lot
8 Expiry date of the test card
3
9 Symbols visible only in GLOW800 mode
9.3

Preparation

WARNING
Surgeon profile without a programmed fluorescence
function
Select the correct surgeon profile.
X
Perform a pre-operative check.
X
Check if microscope illumination lamp is within
X
tolerances (see user manual of ARveo 8).
Please prepare the test as follows:
For general tests use the GLOW800 "_Vascular" or
X
"_Vascular+IGS" preset.
Place the GLOW800 test card below the microscope.
X