6.1

Fluorescence settings

Open the "Surgeon Settings" page.
X
Tap the "Fluorescence Settings" icon.
X
The "Fluorescence Settings" page is displayed:
6.1.1
GLOW800 Brightness (Excitation intensity)
Recommended excitation settings
The default and recommended "Excitation" setting (1) is
100 % to achieve good fluorescence visibility in higher
magnifications and working distances.
Adapt the "Excitation" setting (1) using the slider.
X
Press "Back" successively until you return to the "Options" page.
X
You may be prompted to save the settings.
Save the settings to the profile or choose to only use them once,
X
after which the changes will be discarded.
Press "X".
X
The Live Screen is displayed.
6.1.2
Threshold
With the "Lower Threshold" and the "Upper Threshold", the
fluorescence intensity range to be displayed in the white light
fluorescence view can be defined. Low intensity signals (i.e. noise)
and/or high intensity signals can be filtered out by defining a lower
and an upper threshold with the two sliders. The default values for
the lower and upper thresholds are 0 % and 100 % respectively, to
observe the full range of fluorescence signals.
1
GLOW800 / Ref. 10 748 652 / Version 01
1 2
Increase or decrease the "Upper Threshold" (1) and "Lower
X
Threshold" (2) using the slider bars.
Adjust the "Lower Threshold" values from the live screen using
X
the rotary menu (3).
There is always a difference of 10 % maintained between the
upper and lower threshold values.
Modify the GLOW800 settings
3
13