Safety information
WARNING
CAUTION
2.2
Safety information regarding environmental influences
WARNING
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe – Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Temperature-regulated goods which are stored in the tem-
perature-regulated space when a fault message is displayed
must be inspected for usability before use, because they were
subjected to a temperature higher than 46°C.
Overhaul measures and device modifications may only be
carried out by qualified service personnel or employees of
Barkey GmbH & Co. KG.
The influence of stronger electromagnetic fields (e.g. through
the use of HF therapy or surgical devices) can lead to malfunc-
tions in the device. If such malfunctions occur, increase the
distance between the device and the device causing the dis-
turbance, or operate the devices at separate times. The device
works perfectly within the limit values set in the EN 60601-1-2
standard. The device can be influenced outside the limit values
set by EN 60601-1-2.
Portable and mobile high-frequency communication equipment
such as mobile telephones can influence the device.
The device may not be used in the immediate vicinity of
•
flammable materials (e.g. gases, liquids),
•
flammable mixtures of anaesthetic substances with air,
•
flammable mixtures of anaesthetic substances with oxygen
or laughing gas
whose point of ignition is below 50°C. It is imperative that the
device is not used in areas in which alcoholic disinfectants or
anaesthetising substances are being used simultaneously.
Observe the safety information and the information on electro-
magnetic characteristics and protective distances in the in-
struction manual.
The device may not be set up or operated in the immediate
vicinity of devices with high heat output.
Service manual
Barkey warming center III DTM
5