Replacing The Power Switch - Barkey warming center III DTM Service Manual

With digital temperature display
Table of Contents

Advertisement

Service manual
Barkey warming center III DTM
7.7

Replacing the power switch

Removal:
Remove the heater insert (see Chapter 7.2).
Remove the cover plate (see Chapter 7.2).
Pull the spade connectors off the power switch.
Press the power switch forwards out of the front panel.
Refitting the power switch:
Refitting the power switch takes place in reverse order to removal:
Press the power switch into the front panel cutout until it engages.
NOTE
The label for the switched on condition (" I ") must be facing upwards when fitted.
blue
Figure 33: Power switch connections
Refix the cover plate (see Chapter 7.2).
Refit the heater insert (see Chapter 7.2).
32
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
brown
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Push the spade connectors onto the
power switch:
the cable from the fuse goes to
the bottom contact
the cables from the controller cir-
cuit board go to the top contact