Replacing The Mechanical Overtemperature Protection - Barkey warming center III DTM Service Manual

With digital temperature display
Table of Contents

Advertisement

Service manual
Barkey warming center III DTM
7.6

Replacing the mechanical overtemperature protection

Removal:
Remove the heater insert (see Chapter 7.2).
Remove the cover plate (see Chapter 7.2).
Figure 29: Heater unit fixing screws
Figure 30: Overtemperature protection fixing screw
Remove the mechanical overtemperature protection from the heater unit.
Pull both connecting cables off the overtemperature protection.
Unscrew the cylindrical metal adapter from the overtemperature protection.
Refitting the overtemperature protection:
Refitting the overtemperature protection takes place in reverse order to removal:
Screw the cylindrical metal adapter to the overtemperature protection.
30
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Loosen the four heater unit screws
and pull the heater unit out towards
the front.
Loosen the screw which fixes the
overtemperature protection to the
heater unit.
Service work