Adjustments
6
Adjustments
If there are discrepancies during the temperature check or during testing of the overtem-
perature protection, the Barkey warming center III DTM must be readjusted.
WARNING
The adjustment partly takes place to live components when the device is open.
Adjustments may only be carried out by qualified service personnel and employees
of Barkey GmbH & Co. KG.
In order to carry out adjustments, you must remove all temperature-regulated
goods from the heating chamber and store them in accordance with their respec-
tive manufacturer's instructions.
6.1
Adjusting the temperature display
If the temperature displayed on the Barkey warming center III DTM temperature display
does not match the temperature in the heating chamber, the temperature display needs
to be adjusted.
►
Check the temperature in the heating chamber and the temperature indicated on the
temperature display as described in Chapter 5.2.
►
Switch the device off, and remove the mains cable.
Figure 8: Regulator circuit board cover panel
Figure 9: Regulator circuit board
►
Reattach the cover on the Barkey warming center III DTM insert.
►
Reconnect the mains cable and switch the Barkey warming center III DTM on.
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe – Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Barkey warming center III DTM
►
Remove the cover on the Barkey
warming center III DTM insert.
►
Correct the temperature displayed on
the device using trimmer P4 on the
regulator circuit board. Here, one
revolution of the setting screw corre-
sponds to around 1°C.
If the screw is rotated clockwise, the
displayed value will be reduced.
Service manual
17