Safety Information; Safety Information For The Carrying Out Of Service Work - Barkey warming center III DTM Service Manual

With digital temperature display
Table of Contents

Advertisement

Service manual
Barkey warming center III DTM
2

Safety information

2.1

Safety information for the carrying out of service work

WARNING
4
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
Read this service manual and the instruction manual thor-
oughly through before carrying out a TSI or service work.
Observe the safety information and the information on electro-
magnetic characteristics and protective distances in the in-
struction manual.
All electrical installations must comply with the relevant appli-
cable regulations and standards in addition to the specifica-
tions stated by the manufacturer.
Only mains power cables supplied by Barkey GmbH & Co. KG
which are designed for the device rated voltage and rated cur-
rent may be used.
You must remove the plug from the mains before opening the
device.
A technical safety inspection must be carried out annually by
qualified service personnel or employees of Barkey GmbH &
Co. KG.
All inspection steps for the TSI which have to do with tem-
perature measurements must be carried out at a room tem-
perature of
22° Celsius ± 1.0 K.
During measurements, the device may not be subjected to
direct sunlight or draughts, or any other influences which might
influence the measurement.
The heating device must be located on a level, horizontal sur-
face. When the device is in location, the front castors must be
locked.
Before the heating device is moved, the mains power cable
must be unplugged from the wall socket.
Moving the heating device may only take place slowly and
without temperature-regulated goods at a pace slower than
walking pace. If moving takes place too quickly, or abrupt ac-
celeration or braking is used, the doors can open unintention-
ally despite the closing technology used (magnetic sealing
frame).
Only move the heating device across level surfaces. Do not
move it across steps, inclines or cables etc.
If the temperature display is blinking, the heating device may
not be used for the warming of heating goods.
If the fault display illuminates or flashes, the heating device
must be inspected in the manufacturer's works or by medical
technicians who have been trained in the overhaul and exami-
nation of this device, and overhauled if necessary.
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Safety information

Advertisement

Table of Contents
loading

Table of Contents