Replacing The Transverse Airflow Fan - Barkey warming center III DTM Service Manual

With digital temperature display
Table of Contents

Advertisement

Service work
NOTE
Ensure that there is a spacer roll on each screw.
Strip the insulation off all four cables for around 8 mm. Crimp all four heating rod ca-
bles into one cable end sleeve. Now crimp the other three cables together in one ca-
ble end sleeve.
Figure 22: Heater connection
Refix the cover plate (see Chapter 7.2).
Refit the heater insert (see Chapter 7.2).
7.5

Replacing the transverse airflow fan

Removal:
Remove the heater insert (see Chapter 7.2).
Remove the cover plate (see Chapter 7.2).
Figure 23: Fan connections
Cut the three cables which are fixed in a common cable end sleeve directly at the ca-
ble end sleeve.
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe – Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Barkey warming center III DTM
Reconnect both cables to the regula-
tor circuit board pin-and-socket con-
nector.
The three cables are connected un-
derneath.
Loosen the clamping screws on the
two fan cables at the pin-and-socket
connectors on the regulator circuit
board, and remove the cables.