Technical safety inspections (TSI)
Allow the Barkey warming center III DTM to cool down for 30 minutes with the door
open.
Reset the triggered mechanical overtemperature protection.
Figure 6: Reset pin for mechanical overtemperature protection
5.4

Measurements in accordance with DIN EN 60601-1

Safety inspections must be carried out in accordance with DIN EN 60601-1 in its valid
edition. When carrying out the inspection, you must also read and observe the test
equipment operating manual.
Enter the type and serial number of the test equipment in the inspection log.
Connect the Barkey warming center mains cable to your test equipment.
NOTE
The power switch must be switched on during the test (position "I").
Carry out all measurements in accordance with your test equipment operating man-
ual.
Figure 7: Measuring points for safety check
Enter the values determined for the earth conductor resistance, insulation resistance,
earth and housing leakage current in the inspection log.
Carry out an evaluation based on the measured results, and taking into account pre-
vious electrical safety measurements, to determine whether the device meets the
safety requirements.
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe – Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Barkey warming center III DTM
The overtemperature protection reset pin
is located on the underside of the heater
inset. Press the reset pin in with a pointed
object such as a ballpoint pen about 1
mm.
In the adjacent figure you can see the
measuring points which the test equip-
ment probe must be connected to for
measuring the protective earth and
ground conductor resistance.
Service manual
15