Wait around 60 minutes so that the temperature in the Barkey warming center III
DTM can adjust to the new adjusted settings.
Check the temperature in the heating chamber and the temperature indicated on the
temperature display as described in Chapter 5.2.
Repeat the temperature display adjustment until the temperature display varies by a
maximum of ± 1°C from the reference thermometer.
You must now carry out a technical safety inspection. This process is described in
Chapter 5.
6.2

Adjusting the temperature in the heating chamber

If the temperature measured in the Barkey warming center III DTM is not 42°C ± 3°C, the
temperature setting needs to be corrected.
Check the temperature in the heating chamber and the temperature indicated on the
temperature display as described in Chapter 5.2.
Switch the device off, and remove the mains cable.
Figure 10: Regulator circuit board cover panel
Figure 11: Regulator circuit board
Reattach the cover on the Barkey warming center III DTM insert.
Reconnect the mains cable and switch the Barkey warming center III DTM on.
Wait around 60 minutes so that the temperature in the Barkey warming center III
DTM can adjust to the new adjusted settings.
Check the temperature in the heating chamber and the temperature indicated on the
temperature display as described in Chapter 5.2.
Repeat the temperature adjustment until the temperature in the heating chamber is
42°C ± 3°C.
18
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Remove the cover on the Barkey
warming center III DTM insert.
Correct the temperature setting using
trimmer P2 on the regulator circuit
board. Here, one revolution of the
setting screw corresponds to around
2°C.
If the screw is rotated clockwise, the
displayed value will be increased.
Adjustments