Functional inspection

Switch the Barkey warming center III DTM on.
The green LED on the display panel illuminates to confirm that the device has been
switched on. Enter whether the green LED illuminates in the inspection log.
The temperature of the Barkey warming center III DTM must be displayed in the tem-
perature display after switching on. The temperature display should not be blinking.
Enter whether the temperature display is permanently illuminated in the inspection
log.
Open the cabinet door.
If the door is open, the green LED in the display panel blinks. In addition, an acoustic
signal is given off after 2 minutes to remind you that the cabinet door is open. Enter
whether the green LED is blinking in the inspection log.
Enter whether an acoustic signal was sounded after 2 minutes in the inspection log.
Close the cabinet door.
The acoustic signal will stop and the green LED on the display panel will be perma-
nently illuminated. The circulating air temperature regulation will be activated at the
same time. This will be audible due to the ventilator noise. Enter whether the circu-
lating air temperature regulation is functioning in the inspection log.
5.6

Comments

This is for entering comments, for example information relevant to subsequent tests.
5.7

TSI has been passed

If all the points for the TSI listed above were completed positively, then the TSI in-
spection has been passed. Please tick the box marked "yes".
5.8

Affixing the test sticker

If the TSI test has been passed, the test sticker must be affixed. The test sticker shows
the user the date for the next TSI.
Attach the test sticker with the date of the next inspection to the Barkey warming
center III DTM. The best location to this is the left-hand side of the incline on the
heater insert. The label should be fixed so that the arrow is pointing to the month in
which the inspection is due.
The name and address of those carrying out the test must be entered in the inspection
log. Correct implementation of the TSI inspection must be confirmed in the inspection log
with statement of name, date and signature.
WARNING
The Barkey warming center III DTM must be allowed to cool down to a temperature
below 42°C in the heating chamber after the TSI before it is used for warming
again.