Replacing the microswitch

Removal:
Remove the heater insert (see Chapter 7.2).
Remove the cover plate (see Chapter 7.2).
Undo the nuts from the microswitch using a 14mm jaw spanner, and pull it off the
front panel to the rear.
Pull the spade connectors off the microswitch.
Refitting the microswitch:
Refitting the microswitch takes place in reverse order to removal:
Press the microswitch into the cutout in the front panel, and tighten the nuts using a
14 mm jaw spanner.
NOTE
The microswitch must be flush with the front panel from behind so that it switches
correctly when the door is closed.
Figure 35: Microswitch collections
Refix the cover plate (see Chapter 7.2).
Refit the heater insert (see Chapter 7.2).
34
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: info@barkey.de
Push the spade connectors onto the
microswitch:
one cable to the middle contact
the other cable to the rear contact
The polarity is not important.