Radius Correction - Hermle Z 207 A Operating Manual

Small centrifuge
Table of Contents

Advertisement

Figure 13
This table shows the acceleration and deceleration stages 0 and 9:
Rotor-Number
221.57 V01

2.5 Radius Correction

If adapters or reducers are being used, it could change the centrifugal radius of the respective rotor.
In that case, the radius can be corrected manually.
Please proceed as follows:
Press the key "time" (6) and the key "prog" (11), at the same time and hold down. In the display,
"time" (A-3), appears the word "radius" (M9). With the control field (1), pre-select the respective
radius correction, (see Table 3, APPENDIX P. IV), in steps of 0.1cm.
As soon as the radius correction is set, the word "radius" (M9) appears. This text remains visible
until the radius correction is set back to 0.
For all number marked text, please refer to Figure 14.
Figure 14
© Hermle Labortechnik GmbH
Acceleration
Stage 0
Stage 9
35
8
in seconds
Acceleration time
-1
from 0 min
- n
max
Z207A_V1.17
_eng
Decceleration
Stage 0
Stage 9
25
7
Decceleration time
-1
from n
- 0 min
max
OPERATING
15