Determining the lamp type

Note: This feature is not included on the MiniRAE Lite+.
The monitor may be able to accommodate three lamp values: 9.8eV, 10.6eV (standard) and
11.7eV.
There are two ways to determine the lamp type.
1. Turn off the instrument and remove the lamp. Now look at the serial number. The following
identify the lamp type:
10.6eV: SN:106 XXXXXXX
l
9.8eV: SN:098 XXXXXXX
l
11.7eV: SN:117 XXXXXXX (available for UltraRAE 3000+ and MiniRAE 3000+)
l
2. When the monitor is running, the lamp type is shown along with the calibration and
measurement gas and Correction Factor:
3. Note: This screen can be accessed from the reading screen by pressing [N/-] four times.
When using an11.7eV lamp, refer to TN-207 for the calibration process.
Programming the lamp ID
The correct measurement gas library is used by the instrument when you ensure that the right
lamp value is programmed.
To manually select the Lamp ID:
1. Enter the Programming menu.
2. Select Monitor Setup.
3. Scroll down and select the Lamp ID sub-menu.
4. Press [N/-] to scroll down to the desired Lamp ID.
5. Press [Y/+] to select.
6. Press [MODE] to select Done.
7. Select "Save."
8. Return to the main menu.
Recalibrate the instrument before returning it to service.