Monitor setup

Many settings can be accessed in this menu, including setting the date and time and adjusting
the pump's on/off duty cycle.
MiniRAE 3000+, ppbRAE 3000+, UltraRAE 3000+
Radio power
The radio connection can be turned on or off. (The default value is off.)
1. Press [N/-] to step from one option to the next (on or off).
2. Press [Y/+] to make your selection (the dark circle in the "radio button" indicates that the
option is selected).
3. When you have completed your selection, press [MODE].
Press [Y/+] to accept the new radio setting (on or off).
l
Press [N/-] to discard the change and move to the next sub-menu.
l
Op mode
Note: This feature is not included on MiniRAE Lite+.
Under Monitor Setup is "Op Mode."
Press [Y/+] to select.
You see two options (one is highlighted):
Hygiene
Search
The current mode is indicated by a dark circle within the circle in front of either Hygiene or
Search.
1. Select Hygiene or Search by pressing [N/-]. The highlighting changes from one to the other
each time you press [N/-].
2. Press [Y/+] to select that mode for the instrument.
3. Press [MODE] when you want to register your selection to place the instrument in the
selected mode.
4. Press [Y/+] to commit the change and exit to the Monitor Setup screen, or press [N/-] to
Undo (exit to the Monitor Setup screen without changing the Mode).