Advertisement

Response time (T
):
90
Accuracy (Isobutylene): 3% at calibration point
PID Detector:
Correction Factors:
Calibration:
Calibration reference:
Inlet Probe:
Radio module:
Wireless Network:
Wireless Frequency:
Modulation:
RF Power (Tx):
Keypad:
Direct Readout:
Intrinsic Safety:
EM Interference:
Alarm Setting:
Operating Mode:
Alarm:
Alarm Type:
Real-time Clock:
Datalogging:
Communication:
Sampling Pump:
Handheld PID Monitors
2 seconds
Easy access to lamp and sensor for cleaning and replacement
Over 200 VOC gases built in
Two-point field calibration of zero and standard reference gases
Patented Reflex PID™ technology
Store up to 8 sets of calibration data, alarm limits, and span values
Flexible 5" tubing (a short tube is also available)
Separation tube housing with permanent VOC tube
BLE (2.4GHz), Bluetooth (2.4GHz) or RF module (868MHz or 915MHz)
Mesh RAE Systems Dedicated Wireless Network (or Wi-Fi/LoRa network for Wi-Fi/LoRa-
equipped instruments)
Bluetooth Low Energy (BLE) module
ISM license-free band, 902 to 907.5 MHz and 915 to 928 MHz, FCC Part 15, CE RE-D
802.15.4 DSSS BPSK
10dBm
1 operation key and 2 programming keys; 1 flashlight switch
Instantaneous, average, STEL, TWA and peak value, and battery voltage
ATEX
IECEx
EX ia IIC T4 Gb (with bat pack 059-3051-
000)
EX ia IIB T4 Gb (with bat pack 059-3052-
000)
CSAE 21UKEX2051X
Highly resistant to EMI/RFI. Compliant with EMC RE-D (RF Modules)
Separate alarm limit settings for Low, High, STEL, and TWA alarm
Hygiene or Search mode
Buzzer 95dB at 12" (30cm) and flashing red LEDs to indicate exceeded preset limits, low
battery voltage, or sensor failure
Latching or automatic reset
Automatic date and time stamps on datalogged information
260,000 points with time stamp, serial number, user ID, site ID, etc.
Upload data to PC and download instrument setup from PC via USB on charging station.
Enhanced datalogging capability through BLE module and mobile app
Internally integrated. Flow rate: 450 to 550 cc/min.
2460
                                                 Ex ia IIB T4 Gb
Sira 17ATEX2082X
CSA 10.0005X Ex ia IIC T4 Gb
                             Ex ia IIB T4 Gb
122
II 2G Ex ia IIC T4 Gb