6 Instruction for Use of Pump Head

6.1 Description and Preparation

6.1.1 Description
The LPP-200 Multi-Use pump heads are designed for
applications in the Life Science industries operating with
®
the PuraLev
i100MU pump system. The wetted parts are
made out of biocompatible materials (FDA, USP VI,
Animal-Free). The pump head is mounted to the
Bearingless Motor (a MagLev drive motor) and the motor
is connected to a controller generating the currents for
rotation and the levitation of the impeller.
Pump Housing
Bottom
4 Screws M3x14mm
(For fixation of
Pump head on motor)
Impeller
Pump Housing
Lid
Inlet
6 Screws M4x15mm
(fixation of pump housing)
Figure 33: Pump head LPP-200.17 with driver IPD-100
Before starting using the LPP-200 Multi-Use pump heads
make yourself familiar with the following warnings,
cautions and instruction.
6.1.2 Inspection Prior to Use
The pump head should be inspected prior to use for any
damage. Do not use the pump head if any damage is
found. Contact Levitronix
suspected pump head.
6.1.3 Traceability for Troubleshooting
Traceability for trouble shooting is assured with a 6 digit
serial number located on the top of the pump head (see
).
Figure 35

6.2 General Warnings and Cautions

WARNING
!
High Magnetic Field Strength of Pump Impeller
The pump head contains a rotor magnet with high field
strength. Pace makers may be influenced and
magnetic field may lead to contusions. Keep distance
to pace makers and handle pump heads with care.
PL-4075-00, Rev02, DCO# 20-275
Driver IPD-100
Outlet
Drain Port
(Option without Drain Port)
Triclamp Sealing
Gasket
®
regarding return of any
User Manual for PuraLev
CAUTION
Magnetic Forces
Pay attention to the magnetic forces when handling the
pump head. It has to be avoided that magnetic parts
are
attracted,
which
contamination. Specifically pay attention to the
magnetic forces, when handling two pump heads at the
same time.
CAUTION
!
The rotating impeller could cause injury. Do not run
the pump system when opening the pump head.

6.3 Replacement of Pump Head

6.3.1 Preparation
After removal of a new pump head from its packaging
assure that no metallic part is magnetically attaching to
the pump head. Before insertion into the motor remove
the Fixation Disk, which comes delivered with the
packaging in order to magnetically fix the impeller against
movements during transportation.
Figure 34: Metallic Fixation Disc
6.3.2 Power Down the System
Set the speed to 0 rpm, disable the system. After running
at higher motor or liquid temperature the pump head
might stick to the motor due to thermal expansion effects.
In this case let the system cool down for a while before
starting the replacement procedure.
6.3.3 Pump head Removal from Motor
Un-tighten the screws M1 to M4, which fix the pump head
to the motor (see Figure 35). Do not use the pump
housing screws P1 to P6. Remove the pump head axially
and attach the Fixation Disk from the new pump head to
it.
®
i100MU
www.levitronix.com
can
cause
cracks
or
37