Electrical Installation Of Easyconnect Model - Levitronix PuraLev i100MU User Manual

Hide thumbs Also See for PuraLev i100MU:
Table of Contents

Advertisement

4.2 Electrical Installation of EasyConnect Model

4.2.1 Overview of Connections and Designations
Fieldbus IN
Fieldbus OUT
Figure 24: Electrical connections to EasyConnect driver IPD-100.4
Pin
Wire
Pin #
1
Name
Color
B
P +
red
A
P -
black
C
NC
--
1-14
NC
--
1: Wire colors of compatible cables: ICP-2.1-xx.
Pin
Wire
Pin #
1
Name
Color
brown
3
GND
black
shield
6
NC
pink-blue
1
RS485+
white
2
RS485-
yellow
4
Internal
grey
5
Internal
red
Pin
Wire
Pin #
1
Name
Color
brown
3
GND
black
shield
6
Pout
pink-blue
1
RS485+
white
2
RS485-
yellow
4
Internal
grey
5
Internal
red
PL-4075-00, Rev02, DCO# 20-275
Standard
Description
Designation
+ 24 VDC
Power Supply
Ground/Earth
Not connected
--
Not connected
--
Table 9: Power supply connector of EasyConnect driver
2: Designations for standard firmware. For other firmware refer to relevant documentation.
Standard
Description
Designation
Ground
--
Not connected
--
RS485+
Fieldbus
RS485-
Internal
Do not
connect.
Internal
Table 10: Fieldbus IN connector of EasyConnect driver
1: Wires of compatible cables: ICS-1.1-xx and ICS-1.2-xx
Standard
Description
Designation
Ground
Supply
Output
Output 24VDC
RS485+
Fieldbus
RS485-
Internal
Do not
connect.
Internal
Table 11: Fieldbus OUT connector of EasyConnect driver
1: Wires of compatible cables: ICS-1.1-xx and ICS-1.2-xx
Specifications
2
Typical Levels
Voltage: 24 VDC ±10%
Power: 110 W
Wire of compatible cables is cut.
Wire of compatible cables is cut.
Typical Levels
--
--
Modbus protocol.
--
Typical Levels
24VDC 10%, 200 mA
(Max. current is together with Pout of PLC
connector in Table 12)
Modbus protocol.
--
User Manual for PuraLev
www.levitronix.com
PLC Connector
Power Supply
Connector
Note
"P -" shall be connected to earth
Wire of compatible cables is cut.
Wire of compatible cables is cut.
Note
--
--
For remote control with other devices.
®
Usage of Levitronix
Service Software with a
RS485 to USB adaptor cable.
Internal bus needed to connect pumps for serial
pumping.
Note
For supply of external devices (displays etc.).
Internally protected.
For remote control with other devices.
®
Usage of Levitronix
Service Software with a
RS485 to USB adaptor cable.
Feed through for multi-pump arrays ( for
example serial pumping, see Figure 6)
Internal bus needed to connect pumps for serial
pumping.
®